EPIC II CRITICAL CARE BED (CONTINUED)
Recommended separation distances between portable and mobile RF
communications equipment and the Epic II Critical Care Bed.
The
Bed is intended for use in an electromagnetic environment in which radiated RF
Epic II Critical Care
disturbances are controlled. The customer or the user of the
electromagnetic interferences by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the
to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Return To Table of Contents
2-30
EMC Information
Epic II Critical Care
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d=1,2
0,12
0,38
1,2
3,8
12
2031-009-011 REV B
Bed can help prevent
Epic II Critical Care
Bed as recommended below, according
m
800 MHz to 2,5 GHz
d=1,2
0,12
0,38
1,2
3,8
12
d=2,3
0,23
0,73
2,3
7,3
23