7.3 Foot switch settings

After connecting the foot switch at the back, the required mode must be set at the touchscreen. To set
the mode, touch the field indicating the foot switch symbol. This changes the text which appears
above the symbol.

7.4 Handles with buttons

Activating the handle with buttons takes place independently of the foot switch settings. These cannot
be activated using the foot switch.

7.5 Neutral electrode settings and attachment

In the monopolar operating mode, a neutral electrode is required.
After attaching the neutral electrode to the patient, the HF unit recognizes the neutral
electrode automatically (cf. illustration on the left and chapter 4.2 Overview of symbols
for operating modes and settings). The HBS x-touch 200i recognizes neutral electrodes
with divided and undivided contact surfaces.
!
If the displayed neutral electrode type does not agree with the used
neutral electrode type, this is indicative of poor adhesion of the
neutral electrode on the patient, a defective neutral electrode or a
Please note the following information when attaching the
All manuals and user guides at all-guides.com
defective cable!
neutral electrode