5 Specifications
Mains connection
Supply voltage
Nominal frequency
Output power in monopolar operation
Pure cut
Blend cut
Polypectomy Mode
Monopolar Cut in moist environment (TUR)
Soft coagulation
Pure coagulation
Spray coagulation
Monopolar coagulation in moist environment (TUR)
Output power in bipolar operation
Bipolar Cut
Bipolar Cut in moist environment (TUR)
Bipolar coagulation (Mode „Bipolar Forceps")
Bipolar coagulation in moist environment (TUR)
Bipolar coagulation (Mode „Bipolar Forceps AUTO
38
All manuals and user guides at all-guides.com
Bipolar AutoStop function permits automatic interruption
of the energy supply when a defined tissue resistance
„Exit": saves the setting and return to the main menu.
Status
MONO
BI
is reached
Meaning
Undivided neutral electrode detected
Divided neutral electrode detected
Neutral electrode not detected or not
connected
The foot switch is set to MONOPOLAR
The foot switch is set to BIPOLAR
100-260 Volt, 50 - 60 Hz.
500 kHz
200 W at 500 Ω
200 W at 500 Ω
80 W at 250 Ω - 160 W at 300 Ω (0,1s)
200 W at 500 Ω
140 W at 500 Ω
140 W at 500 Ω
60 W at 1500 Ω
140 W at 500 Ω
140 W at 250 Ω
140 W at 100 Ω
100 W at 100 Ω
140 W at 100 Ω
100 W at 100 Ω
HBS x-touch 200i