Continuity Measurements; Diode Measurements - Sonel CMP-400 Manual De Uso

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

5.4

Continuity Measurements

Measurements must not be realised in live circuits. Capaci-
tors must be discharged.
Do not realise measurements if the battery compartment is
open.
In order to realise continuity test it is necessary to realise the
following actions:
place the rotational selector in the position Ω
connect the red test lead to sockets V/Ω/TEMP and the black
one to sockets COM,
press the MODE button until
make sure symbol
the case of shorted terminals of the measurement probes the
display shows „0",
touch the test probe tips to the circuit under test,
read the result of the measurement on the display; if the re-
sistance is < 50Ω, a tone will sound,
having done the test disconnect the test leads from the meter.
5.5

Diode Measurements

Measurements must not be realised in live circuits. Capaci-
tors must be discharged.
42
WARNING:
WARNING:
is displayed with open test leads, and in
WARNING:
CMP-400 – USER MANUAL
appears in the display,
,