Xylem SI Analytics CT 72/P Manual De Instrucciones página 73

Termostatos transparentes
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

7.1.
MENU PROGRAM – START
Start-Menu
Menu level 1
Level 2
Menu functions
This menu will start a previously set program.
Requirements:
1. Create a program. (refer to next chapter)
2. Return to the Start-MENU and confirm the desired setting of each MENU
item with the key
3. Set a start time (>TIME< >DATE< >YEAR<) if the program is to be started
by the internal timer.
> STEP<
> RUNS <
> END<
> GO <
Parameter level
or
or
Program start at section 1 ... 10
Number of repetitions 1 ... 99
Status at end of program (STDBY/SETPNT)
Standby or last setpoint
Time of start (NOW/TIMER)
Press the
key if a parameter is to be retained.
Correction function for parameters or values (prior to
ENTER)
Set program step with
and
example: STEP 1
Set number of runs with
and
example: 1 run
Set desired parameters with
and
(STDBY / SETPNT)
Parameter STanDBY: the circulator switches to –OFF-.
Parameter SETPoiNT: the circulator constantly keeps
the temperature at the value of the last step.
Confirm >NOW< with the
start immediately
or
start at the set time under parameter (TIMER ).
Set time in the example below:
09. August 2009, 11:15 hrs
73
.
key and the program will
Tabla de contenido
loading

Tabla de contenido