10.4. Messages de status

Messages de status
00 MANUAL STOP
01 MANUAL START
02 REMOTE STOP
04 REMOTE START

10.5. Messages de panne

Messages de panne
-01 TEMP / LEVEL ALARM
-02 REFRIGERATOR ALARM
-03 EXCESS TEMPERATURE WARNING
-04 LOW TEMPERATURE WARNING
-05 TEMPERATURE MEASUREMENT
ALARM
-06 SENSOR DIFFERENCE ALARM
2
-07 I
C-BUS ERROR
-08 INVALID COMMAND
-09 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
-10 VALUE TOO SMALL
-11 VALUE TOO LARGE
-12 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS
-14 EXCESS TEMPERATURE
PROTECTOR ALARM
-30 CONFIGURATION ERROR: CONFIRM
BY PRESSING <OK> ON
CIRCULATOR
-33 SAFETY SENSOR ALARM
Commande à distance
Description
Thermostat en mode "OFF".
Thermostat en commande manuelle (par le clavier).
Thermostat en mode "r OFF".
Thermostat en commande à distance (par ordinateur).
Description
Alarme sous-niveau.
Court circuit ou coupure dans le cable de commande du
compresseur.
Alarme de la limite de température haute
Alarme de la limite de température basse
Court circuit ou coupure de la sonde interne.
Alarme différence de T° entre sondes.
Sonde de travail et sonde de sécurité ont plus de 35 °C de
différence.
Panne interne lecture ou écriture du I
Ordre inconnu.
Cet ordre n'est pas autorisé dans ce mode.
La valeur entrée est trop petite.
La valeur entrée est trop grande.
Panne dans le convertisseur A/D.
Alarme température de sécurité.
La configuration du thermostat ne correspond pas à son
emploi actuel.
Appuyez sur Enter
uniquement une fois, la configuration.
Le câble de la sonde de sécurité de surchauffe est coupé ou
en court circuit.
143
2
C-Bus.
pour modifier automatiquement,