1.

Technical data of Transparent Thermostats CT 72/P

Translation of the legally binding German version
(Version as of 10. May 2010)
Country of origin: Germany / Made in Germany
1.1.
Technical data of immersion thermostat CT 72
Temperature ranges:
Operating temperature range
with additional cooling
Working temperature range
(20 °C ambient temperature)
with tap water cooling*
with flow-through cooler CK300/CK310 (option)
Temperature stability
Temperature drift at changing ambient temperature
between 5 and 40 °C:
Temperature control
Heater capacity:
Pump :
Connections:
electrical:
connector:
control cable to the flow-through cooler
CK300/CK310
Computer interface
mechanical
nozzles for flow-through cooler. CK300 / CK310
nozzles for tap water cooling
Overall dimensions
Weight
Mains power connection
Current draw
Mains power connection
Current draw
* If tap water is used, an operating temperature of approx. 3 °C above the given tap water temperature can be
.
maintained
Technical data of Transparent Thermostats CT 72/P
at 230 V
at 115 V
Pressure max. at 0 liter
Flow rate at 0 bar
(WxDxH)
230 V/50-60 Hz
(at 230 V)
115 V/60 Hz
(at 115 V)
°C
+ 10 ... + 60
°C
+ 30 ... + 60
°C
+ 20 ... + 60
°C
+ 10 ... + 60
K
± 0,02
max. 0,002 K/1°C
PID3 cascade temperature control
W
2000
W
1000
bar
0.35
I/min
15
Vdc/mA
24-0 / max. 25
RS232
2 x Ø 12 mm
2 x Ø 8 mm
cm
13x16x33
kg
5.0
V/ Hz
190 ... 253 / 50-60
A
9
V/ Hz
90 ... 127 / 60
A
9
60