A
B
C1
C2
D1
D2
Menu functions
The started program
After the start the program will indicate the currently calculated setpoint in line 2
S XX.XX. The value increases within the time period >TSLICE< until the target
temperature >SETPNT< of the section is reached.
If the time period in a section is set to „0", the next section will not begin until the
target temperature has been reached.
Use the edit keys
every 4 seconds between the current section (STEP XX) and the
A
remaining time of the section
B
remaining time of the program
C
current bath temperature
I xxx.xx - internal actual value or
E xxx.xx – external actual value
D
RUN – the program has started or
PAUSE – the progress of the program has been interrupted by pressing the
key. While the time is stopped the temperature will constantly remain at
the last calculated setpoint
Continue with the
Termination / Interruption of a program
The program can be terminated any time by pressing the key
In case of power failure the program is interrupted.
The circulator switches to –OFF- .
If the AUTOSTART-function is activated the programmer starts again at the point
in time approx. 5 minutes prior to the interruption. However, an uncontrolled
change of the bath temperature has occurred.