7.2.
MENU PROGRAM – creation, administration
Menu level 1
1 program
10 sections
Menu level 1
Level 2
Menu functions
The integrated programmer permits fast and easy programming of setpoint
temperature sequences. This temperature sequence is called program. A program
is composed of individual sections (STEP). The sections are defined by duration
(TSLICE) and target temperature. The target temperature is the setpoint (SETPNT),
which is achieved at the end of a section. The programmer calculates the
temperature ramp from the difference in time and temperature.
STEP
(Nr.)
SETPNT
(°C)
TIME
(hh.mm)
°C
200
150
100
75
50
25
10 20 30 40 50 60 10 20 30 40 50 60 10 [min]
Graph1
Sections without set value and time are skipped. They can be defined
retroactively and the integrated into the program.
>EDIT<
> DELETE<
Press
key, if a parameter is to be retained.
Correction function for parameters or values (prior to ENTER)
Level 3
1
2
100
180
00:20
00.10
Step 1
2
4
Create, administer program
> STEP<
>SETPNT <
>TSLICE<
delete program step (01 ... 10, ALL)
Parameter level
Set program step with
and
 (Example: EDIT STEP 01)
For STEP 01 the values for SETPOINT 01 and
TSLICE 01 are set one after the other
76
4
6
180
75
00:20
00:50
6
8
Program step (1 ... 10)
Temperature setpoint of step ...
Duration of step ...
(STEP 1 ... 10)
8
75
00:20
t