CELL ASSEMBLY REPLACEMENT KIT (CONTINUED)
5 .
Fold back head end of support surface as shown in Figure 3 .
Note:
Two people required to fold back foam crib and head box .
6 .
Cut tubing just above the 90 degree fitting from the head box; two places . Refer to Figure 4 .
7 .
Remove existing 90 degree fittings from head box and replace with new 90 degree fittings, two places . Refer to
Figure 5 for orientation .
Return To Table of Contents
1-50
Service Information
Figure 3
Figure 4
Figure 5
2950-109-005 REV B