Safety check
There are no prescribed or recommended safety checks to be carried out.
Rationales
– The engineering of the Medela Thopaz + has been verified by independent test institutes for
compliance with the EN / IEC 60601-1 standard. Copies of such certificates may be obtained
on request. Because of the method of construction used, Medela does not expect electrical
safety to be affected at any time during the life of the product – provided that Thopaz + is
repaired throughout its service life strictly and exclusively by Medela authorized service centers
and that it is used properly in accordance with the intended use.
– For Thopaz + as a device in protection class II (EN IEC 60601-1), there is no protective earth
conductor; protective earth related tests (protective earth resistance, earth leakage current
etc.) are not applicable.
– Thopaz + enclosure is made entirely of insulating material. Tests of the touch current using
common measuring instruments will therefore not reveal measurable values.
– The applied parts of Thopaz + are connected to Thopaz + via non-conductive vacuum hoses,
canisters and tubings. Even when suctioning a conductive fluid (9 g/L NaCl; EN ISO 10079-
1;1999) until the overflow protection device activates, measurements of the patient leakage
current using common measuring instruments will not reveal measurable values.
– Thopaz + systems do not have patient connections and functional earth conductors; tests
related to those connections and currents are not applicable.
48 |
Warranty, maintenance and checks
1. Connect the tubing
079.0016 to Thopaz + . For guidance to connect the
system refer to chapter 7 – Preparation for use.
2. Connect vacuum gauge with tube to the patient connector.
3.
1
Press simultaneously
079.0021 and the 0.8 l canister
2
3.1 Press and hold [
[
] simultaneously. Press
additionally [
3.2 The safety test starts.
3.3 Follow the instructions on
the display
1
] and
].