10 Changing settings
10.1
Factory settings
Thopaz + is delivered with the following factory settings.
These settings can be changed and saved as the new standard.
Pressure:
-2.0 kPa
– Choose to max. -10 kPa (-100 cmH2O,
-100 mbar, -80 mmHg).
Pressure unit: kPa
– Choose between kPa, cmH2O, mmHg, mbar.
Measured
Hide
pressure:
– Choose between hide and show (only relevant
in the air leak history).
– This feature is added for advanced Thopaz +
users. A higher measured pressure than set
pressure can be an indication that the patient
is able to regulate the negative pressure in the
inter-pleural space.
Fluid time 1:
OFF
– Choose between 24 h / 12 h / 8 h / 4 h / OFF
– Shows the total amount of collected fluid over
the last X hours.
Fluid time 2:
OFF
– Choose between 2 h / 1 h / 0.5 h (30 min) /
0.25 h (15 min)/OFF
– Shows the total amount of collected fluid over
the last X hours/minutes.
Fluid alarm:
OFF
– Choose between 500 ml / 5min, 15min or
60min; 400 ml / 5min, 15min or 60min;
300 ml / 5min, 15min or 60min; 200 ml / 5min,
15min or 60min; 100 ml / 5min, 15min or
60min; 50 ml / 5min, 15min or 60min/OFF
– The Fluid alarm 407 appears if Thopaz +
detected a higher fluid collection in the
last 5 min, 15min or 60min than the desired
parameter.
Air leak
Hide
history:
– Choose between hide and show.
– This history can be displayed on the main
display (see chapter 7.6.1 – Main display).
Language:
English
– Choose your language.
Save as
Choose if you want to save the new parameter
standard:
as a standard (guidance see chapter 10.3 –
Save a changed parameter as the new
standard).
| 39
Changing settings