Table 6-2
Performance verification tests (continued)
Step
Test function
5
Press
to select AC V measurement
6
Turn the rotary switch to the
Press
to select DC mV measurement
Keysight U1253B User's and Service Guide
Range
5 V
[a]
50 V
500 V
1000 V
50 mV
position
500 mV
1000 mV
Performance Tests and Calibration
Reference signals/
values
5 V, 1 kHz
5 V, 10 kHz
4.5 V, 20 kHz
4.5 V, 100 kHz
50 V, 1 kHz
50 V, 10 kHz
45 V, 20 kHz
45 V, 100 kHz
500 V, 1 kHz
1000 V, 1 kHz
50 mV
500 mV
–500 mV
1000 mV
–1000 mV
6
Error limits
± 22.5 mV
± 79.0 mV
± 169.5 mV
± 169.5 mV
± 225 mV
± 790 mV
± 1.695 V
± 1.695 V
± 2.25 V
± 8.0 V
[b]
± 75 V
± 175 V
± 175 V
± 0.75 mV
± 0.75 mV
175