Performance Tests and Calibration

Calibration From Front Panel

Calibration process

The following general procedure is the recommended method to complete a full
instrument calibration.
1 Read and implement
2 Perform the verification tests (refer to
3 Perform the calibration procedures (refer to the
4 Secure the instrument after calibration.
5 Take note of the new security code (if it has been changed) and the calibration
Make sure to quit the adjustment mode before switching off the instrument.
NOTE
192
the instrument.
page 193; see also
"Adjustment Considerations"
count in the instrument maintenance records.
"Test Considerations"
Table 6-2
Keysight U1253B User's and Service Guide
on page 172.
on page 174) to characterize
"Calibration procedures"
on page 187).
on