Performance Tests and Calibration

Valid adjustment reference input values

Adjustments can be performed using the following reference input values:
For serial numbers below MY51530001, the 10 kHz input frequency is applied to
NOTE
those marked with a asterix (*)
Table 6-3
Valid adjustment reference input values
Function
Range
DC mV
Short
50 mV
500 mV
1000 mV
AC mV
50 mV
500 mV
1000 mV
DC V
Short
5 V
50 V
500 V
1000 V
188
Reference input value
SHORT
30.000 mV
300.00 mV
1000.0 mV
3.000 mV (1 kHz)
30.000 mV (1 kHz)
30.000 mV (20 kHz) *
30.00 mV (1 kHz)
300.00 mV (1 kHz)
300.00 mV (20 kHz) *
300.0 mV (1 kHz)
1000.0 mV (1 kHz)
1000.0 mV (20 kHz) *
SHORT
3.0000 V
30.000 V
300.00 V
1000.0 V
Valid range for reference input
Short V and COM terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
Short V and COM terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
Keysight U1253B User's and Service Guide