Immunity test
Conducted RF
3 Veff
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects an people.
Operation and
maintenance manual
Test level to
EN/IEC 60601-1-2
MU_029_EN
Rev.0
Conformity
Electromagnetic environment -
level
Portable
communications equipment should
be used no closer to any part of the
Products, included cables, than the
recommended separation distance
calculated
applicable to the frequency of the
transmitter.
Recommended
distance:
3 Veff
where P is the maximum output
power rating of the transmitter in
3 V/m
watts
transmitter manufacture and d is
the
distance in meters (m).
Field
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance leave
in each frequency range.
Interference may occur
in the vicinity of equipment marked
with the following symbol:
07/11/18
Page 30 of 35
directives
and
mobile
from
the
d = 1,2√P 150 KHz to 80 MHz
d = 1,2√P 80 MHz to 800 MHz
d = 2,3√P 800 MHz to 2.7 GHz