Immunity test
+/- 8 kV at contact
+/- 15 kV in air
Electrostatic
discharge
(ESD)
IEC 61000-4-2
+/- 2 kV
For electric power
Rapid impulse
lines
electric
transistors
+/- 1 kV
For input/output lines
IEC 61000-4-4
+/- 1 kV
Between phases
Overvoltage
+/- 2 kV
IEC 61000-4-5
Between phases and
earth
<5% U
(drop >95% of U
For 0.5 cycles
Voltage dips,
<40% U
short
(drop = 60% of U
interruptions and
For 5 cycles
variations on the
power supply
<70% U
input lines
(drop = 30% of U
For 25 cycles
IEC 61000-4-11
<5% U
(drop >95% of U
For 5 s
30 A/m
Magnetic field at
electrical mains
frequency
(50/60Hz)
IEC 61000-4-8
NOTE U
mains voltage in AC before application of test level.
T
Operation and
maintenance manual
Test level to
EN/IEC 60601-1-2
T
)
T
T
)
T
T
)
T
T
)
T
MU_029_EN
Rev.0
Conformity level
+/- 8 kV at contact
+/- 15 kV in air
+/- 2 kV
For electric power
lines
+/- 1 kV
For input/output lines
+/- 1 kV
Between phases
+/- 2 kV
Between phases and
earth
<5% U
T
(drop >95% of U
)
T
For 0.5 cycles
<40% U
T
(drop = 60% of U
)
T
For 5 cycles
<70% U
T
(drop = 30% of U
)
T
For 25 cycles
<5% U
T
(drop >95% of U
)
T
For 5 s
30 A/m
07/11/18
EN
Page 29 of 35
Electromagnetic
environment - directives
Floors must be made of wood,
concrete or ceramic tiles. If
the floors are covered with
synthetic
material,
humidity must at least be
equal to 30%.
Mains voltage quality should
be that of a typical commercial
or hospital environment.
Mains voltage quality should
be that of a typical commercial
or hospital environment.
Mains voltage quality should
be that of a typical commercial
or hospital environment.
If the Product user requires
continued
function
mains
power
interruptions,
the
should be supplied by a UPS
unit or batteries.
The magnetic fields at mains
frequency should have the
characteristic
levels
typical locality in a commercial
or hospital environment.