MAINTENANCE

6.1

Error message

When the input value/measurement result is out of range or calibration data is not good enough,
the meter will display error information.
If the error cannot be solved by the error message displayed, please contact Ohaus sales
representative for further assistance.
6.2

Meter maintenance

The STARTER 5000 series instruments do not require any maintenance other than occasional
wiping with a damp cloth.
Attention: Follow these safety precautions when cleaning the instrument
Disconnect the instrument from the power supply
Do not open the instrument
Do not allow liquids to enter the instrument
Attention: The housing is made of acrylonitrile butadiene styrene (ABS). This material is attacked
by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage
should be immediately wiped off.
6.3

Electrode maintenance

Make sure the electrode is filled with electrolyte solution. Always store the electrode according to
the electrode instruction manual and do not allow it to dry out.
If the electrode response becomes sluggish or the slope is not good enough, try the following:
o
Soak the electrode in 0.1M HCl for more than 8 hours.
o
For fat or oil contaminant, degrease the membrane with cotton wool soaked in either
acetone or a soap solution.
After electrode treatment, a new calibration should be performed. If the electrode slope is still not
good, the electrode might need to be replaced.