Set up and Commissioning
2.1
Scope of delivery
The scope of delivery is shown on the included packing list.
There is a detailed list of the individual sets in  chapter 11 Scope of delivery
Accessories can be found in  chapter 12 Accessories
2.2
Unpacking and Setup
The meter and all accessories were carefully inspected for function and size accuracy at the factory.
Please make sure that you also remove all of the small additional parts from the packaging.
The meter can be set up on any even surface.
The safety of the system, in which the device will be integrated, is the responsibility of the installer.
If the meter is transported from a cold into a warm environment, there can be malfunctions in the
device caused by condensate. In this case, you have to wait until the device temperature and the
room temperature have reached a balance before you start the meter.
The flawless function and operational safety of the meter can only be warranted under those
climate conditions specified in  1.3 "Specifications".
2.3
Power supply
The meter is delivered ready to operate with a 5 DC power supply.
Position the power supply where it is easily accessible so that the meter can be separated from
the power at any time.
Check the power supply regularly. If the power supply is damaged it must be promptly replaced by an
undamaged unit.
Use only the power supply Z 612 (Mo. VER05US050-JA; input 100-240V~; 50/60Hz; 0,18A;
output 5V DC; 1A max.; class II)!
Applies only to the Lab 745:
After an interruption in the continuous polarization (device separated from power supply or
similar), you must observe the waiting period for calibration. We also recommend a new calibration.