62
9.5
Configuration Lab 845
9.5.1 Basic settings
Via (CONF) you will go to the menu item "Configuration".
The entire structure of the menu (CONF) is displayed graphically in 9.4.
The following device-relevant basic settings can be configured:
data logg.
temperature
Sensortype
general.
9.5.2 Configuration of the fixed temperature
As the pH measured value is always temperature-compensated, the temperature of the measuring
and calibration media must be measured or set up as a fixed temperature.
If the fixed temperature is activated, it is used for temperature compensation.
> The message "FIX" will appear in the top right side of the display.
> Configuration of the fixed temperature:
"Temperature" > "Fixed temperature"
"Temperature" > "Temp. value"
Via (INFO) or (CAL) you select the position to change.
Via (CONF +1) or (LOG -1) you change the position.
(ON/OFF) confirms and saves your setting.
(MEAS) terminates the entry dialog without transferring the value.
9.5.3 Configuration of the sensor type
3 different sensor types can be connected to the Lab 845.
pH
ORP
ISE
Sensor selection:
Select "Sensor type" >, then Type.
Via (CONF +1) or (LOG -1) you selected the desired setting.
(ON/OFF) confirms and saves your setting.
(MEAS) terminates the entry dialog without transferring the value.
Data logger ( Chapter 5)
Temperature
Sensortype
General ( Kapitel 4)
, then select the fixed temperature value.
>
, then select activate / deactivate.
>