Top or bottom bulge in the X-
ray image
X-ray image is mirror-inverted
Round shadow on the X-ray
image
Ghosting or double exposure
on X-ray image
X-ray image mirrored in one
corner
2142100008L41 2202V007
Possible cause
PSP inserted off-center and at
an angle
PSP exposed on the wrong
side.
Plus ID PSP (with marker)
exposed on the wrong side
PSP exposed twice
PSP not sufficiently erased
PSP creased during X-ray expo-
sure
Troubleshooting
Remedy
Check the error code on the
touch screen.
Insert the PSP centrally and
straight.
Insert the PSP correctly in the
barrier envelope.
Position the PSP correctly.
When taking an X-ray, make
sure that the active side faces
towards the X-ray tube.
Only expose the PSP once.
Check the erasure unit for
proper function.
Inform a service technician, if
the problem persists.
Do not bend the PSP.