Check and adjust the specific X-ray unit in accordance with the standard values.
❯
8.5
Commissioning tests
The required tests (e.g. acceptance test) must be done in accordance with local rules and regulations.
Find out which tests are required.
❯
Carry out testing in accordance with local rules and regulations.
❯
Acceptance check
The Intra / Extra Digital test phantom, and possibly the corresponding test phantom holder as
well, is required for acceptance tests with the PSP and sensor as receivers.
Before commissioning the unit, the acceptance test of the X-ray system must be carried out in
❯
accordance with national regulations.
The constancy tests, which must be carried out at regular intervals by the surgery personnel, are
based on the results of the acceptance test.
Electrical safety checks
Carry out the electrical safety check according to national law (e. g. in accordance with IEC 62353).
❯
Document the results.
❯
Carry out and document the instruction and handover for the unit.
❯
A sample handover report is included in the attachment.
2142100008L41 2202V007
without X-ray field
limitation
60 kV
2
mGycm
0.05 s
9.1
0.07 s
12.8
0.11 s
20.1
0.11 s
20.1
without X-ray field
limitation
70 kV
2
mGycm
0.08 s
7.3
0.11 s
10.0
0.14 s
12.8
0.14 s
12.8
DC emitter, 7 mA
Tube length 20 cm
X-ray field limitation
2
0.8 x 1.2 in
2
(20 x 30 mm
)
60 kV
2
mGycm
0.05 s
1.9
0.07 s
2.7
0.11 s
4.2
0.11 s
4.2
DC emitter, 6 mA
Tube length 30 cm
X-ray field limitation
2
0.8 x 1.2 in
2
(20 x 30 mm
)
70 kV
2
mGycm
0.08 s
1.5
0.11 s
2.1
0.14 s
2.7
0.14 s
2.7
Installation
X-ray field limitation
2
1.2 x 1.6 in
2
(30 x 40 mm
)
60 kV
2
mGycm
0.05 s
3.8
0.07 s
5.4
0.11 s
8.5
0.11 s
8.5
X-ray field limitation
2
1.2 x 1.6 in
2
(30 x 40 mm
)
70 kV
2
mGycm
0.08 s
3.1
0.11 s
4.2
0.14 s
5.4
0.14 s
5.4