Check the PSP with a homogeneous test
image
Scratches on the surface of the PSP may be visi-
ble in the X-ray image and may impair the ability
to diagnose the X-ray images. If scratches are
visible in the image that may impair the diagnos-
tics, the PSP must be discarded.
Fig. 8: PSP with scratches
Place the PSP on a level surface at a distance
❯
of approximately 30 cm from the X-ray tube.
Doing this, make sure that the active side of
the PSP faces the X-ray tube.
Set the exposure time and setting values on
❯
the X-ray unit for a molar x-ray image (see "8.4
X-ray unit settings").
Scan the PSP with the PSP scanner at high
❯
resolution.
Check the image for homogeneity.
❯
No scratches visible in the image: PSP can still
be used.
Scratch is visible in the image: discard the PSP.
2142100008L41 2202V007
Check X-ray system
To check the X-ray system, take an X-ray image
with the test phantom (refer to "3.3 Optional
items"). This can be used to check the image
produced with the X-ray system for homogeneity,
resolution, contrast and artifacts.
Take an X-ray image with the test phantom.
❯
Comply with the instructions for use of the test
phantom.
Read the PSP.
❯
Check the image for homogeneity, resolution,
❯
contrast and artifacts.
If errors are visible in the image, contact a serv-
❯
ice technician.
Usage
EN-
US