11 Operation; X-Ray - Air Techniques ScanX Swift View Instrucciones Para Montaje Y Uso

Ocultar thumbs Ver también para ScanX Swift View:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Protect the phosphor storage plates from sun-
light and ultraviolet light.
Store phosphor storage plates in a barrier
envelope or intraoral/extraoral plate protector
of proper size.
Phosphor storage plates will be pre-exposed
on exposure to natural radiation and stray x-ray
radiation. Protect erased or exposed phosphor
storage plates from X-ray beams.
If the phosphor storage plate has been stored
for longer than one week, erase the phosphor
storage plate prior to use.
Do not store phosphor storage plates under
hot or moist conditions. Note the ambient con-
ditions (see "4 Technical data").
When used properly, phosphor storage plates
can be exposed, read and erased several hun-
dred times provided there is no mechanical
damage.
Replace the phosphor storage plate if there are
any signs of damage, e.g. protective layer is
damaged or visible scratches, that could inter-
fere with the diagnosis.
Phosphor storage plates that have a produc-
tion or packaging defect will be replaced by Air
Techniques in the same quantity.
Clean phosphor storage plates properly (see
"12 Cleaning and disinfection").
2142100008L41 2202V007

11 Operation

CAUTION
Image data on the phosphor storage
plate is not permanent
The image data is altered by light, natu-
ral X-ray radiation and scattered X-ray
radiation. This will lead to a reduction in
diagnostic information and clarity.
Read the image data within 30 minutes
of exposure.
Never handle exposed phosphor stor-
age plates without the barrier enve-
lope.
Do not subject an exposed phosphor
storage plate to X-ray radiation before
and during the scanning process. Do
not X-ray during the scanning process
if the unit is in the same room as the X-
ray tube.
Phosphor storage plates must only be
read using a phosphor storage plate
scanner that is approved by Air Techni-
ques.
11.1

X-ray

The procedure is described using a size 2
PSP as an example.
Required accessories:
– PSP
– Barrier envelope of the same size as the PSP
WARNING
Risk of cross contamination when not
using the barrier envelope or when
using the barrier envelope more than
once
Do not use an phosphor storage plate
without a barrier envelope.
Do not re-use the barrier envelope
(disposable item).
Usage
EN-
US
29
Tabla de contenido
loading

Tabla de contenido