Daily Care And Maintenance - Blancone ARCUS+ Instrucciones De Uso

Unidad de activación de blanqueamiento dental led
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 75
7.
To Stop the bleaching device at any time press the START / STOP button. If the device is not
manually stopped it will stop after the selected time has elapsed. Note that the fan will continue
to work for some time in order to cool down the unit.
8.
At the end of the day turn off the device by holding down
the Start / Stop button for 3-4 seconds or by disconnecting it
from the power network.
9. In case of overheating the thermal protection is triggered and
the device will cease to emit light. The display will indicate "Oh"
and the fans will turn on for 1 minute. After cooling, the device
will continue its normal operation, but stays in Pause mode.
Overheating can only happen in an emergency situation or
by accidental clogging of the cooling vents. For this reason
the cooling vents located on the bottom of the bleaching head
must always be unobstructed in order to allow air circulation.
VIII. DAILY CARE AND MAINTENANCE
1.
Disinfection of the PVC window of the bleaching head
For each patient, the emitting window is cleaned
with a cotton swab soaked in an alcohol solution.
2.
Cleaning the device:
To disinfect the device and its parts, spray disinfectant on a piece
of soft cloth or cotton and clean.
Do not use abrasives or solvents, as this may damage parts of the device!
ULTRA
DUETTO
HOME
NIGHT
DUETTO
ACTIVE
CLICK
ARCUS
LIPS
ARCUS
DUETTO
CARE
Dental LED bleaching activation unit - MOBILE
GB
13
Tabla de contenido
loading

Tabla de contenido