Electromagnetic compatibility
The Triangle has been thoroughly tested and respects the limits for electro-
medical devices specified by IEC 60601-1-2 standards. These limits ensure
reasonable protection against hazardous interference in typical medical
installations.
The device generates, uses and radiates radio frequency energy. If not installed
and used according to the instructions in this manual, it may interfere with other
nearby devices. No guarantee is given that interference will not occur under
certain conditions.
This device is suitable for use in professional healthcare facility environments,
i.e. in hospitals, except for near active HF surgical equipment and RF-shielded
rooms of systems for magnetic resonance imaging, where the intensity of
electromagnetic disturbance is high.
Triangle should not be used adjacent to or stacked on other
equipment. If adjacent or stacked use is necessary, Triangle should
be observed to verify its normal operation in the configuration in
which it will be used.
The existence of electromagnetic interference can be verified easily by turning
the device off and back on again. If it is found that the device is indeed interfering
with other equipment, try to solve the problem by adopting one of the following
solutions:
-
Change the orientation and/or position of the affected device.
-
Move the two devices further away from each other.
-
Contact the manufacturer or authorised service center for further
assistance.
List of cables, transducers and accessories
Cables, transducers and accessories for which Inventis claims compliance with
the IEC 60601-1-2 standard are those supplied with the device itself, as specified
in section 2.2.