Storz ENDOMAT UP 210 Manual De Instrucciones página 66

Ocultar thumbs Ver también para ENDOMAT UP 210:
Tabla de contenido
58
Technische Beschreibung
Brief description (1 line with approx. 40 characters possible)
Possible cause
259: Electronics error
BE: Internal error
300: High pressure
Stopcock on instrument closed
301: Maximum pressure
Excess pressure alarm in URO and HYS application if 150 mmHg is exceeded
500: Main functions not active
FE: A touch screen activation lasting more than 25 s has been detected
501: Touch calibration
FE: A touch screen activation lasting more than 30 s has been detected
0x750: Defective electronics
Electronics error in the cartridge module
0x760: Defective drive
Electronics error in the drive control unit
0x761: Defective control unit
Electronics error in the general control unit
0x762: Defective H bridge
Electronics error in the H bridge
0x763: Motor current offset
Electronics error in the motor current measurement
0x764: Peak current
Electronics error, overcurrent
0x765: Index missing
Electronics error in the speedometer
0x766: Index moved
Electronics error in the speedometer
0x767: Motor PWM
Control error in the motor
Technical description
Information messages
Response
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Observe surgical field, ensure pressure reduction
Observe surgical field, ensure pressure reduction
Acoustic and visual signal – device can still be operated:
A touchscreen calibration can be started by pressing the touchscreen for 25 seconds. This
text message will appear after 25 seconds.
Remedy: Check touchscreen and clean if necessary. This text message may also appear in
the event of a touchscreen short circuit. In this case: Turn the device off and on again, and
contact KARL STORZ Service if the error occurs again.
This text message is part of the touchscreen calibration procedure (see text message 500)
and is displayed after the touchscreen is touched continuously for 30 seconds.
The screen must be released and touched again within 5 seconds in order to start
calibration.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Descripciones técnicas
Tabla de contenido
loading

Tabla de contenido