be sure to keep the i600 device at room temperature for at least
two hours to prevent condensation. If condensation is visible on
the product surface, the i600 should be left at room temperature
for more than 8 hours.
▪......
... Y ou should only disconnect the i600 system from the power supply
via its power cord.
▪......
... W hen disconnecting the power cord, hold the plug surface to
remove it.
▪......
... T he EMISSIONS characteristics of this equipment make it suitable
for use in industrial areas and hospitals (CISPR 11 Class A). If it
is used in a residential environment (for which CISPR 11 Class B
is normally required) this equipment might not offer adequate
protection to radio-frequency communication services.
▪......
... B efore disconnecting the power cord, make sure to turn off the
power on the device using the power switch on the main unit.
▪......
... O nly use the power adaptor supplied together with the i600. The
use of other power adaptors may result in damage to the system.
▪......
... A void pulling on the communication cables, power cables, etc.
used with the i600 system.
5.6 Eye Safety
WARNING
▪......
... T he i600 system projects a bright light from its tip during
scanning.
▪......
... T he bright light projected from the tip of the i600 is not harmful
to the eyes. However, you should not look directly at the bright
light nor aim the light beam into the eyes of others. Generally,
intense light sources can cause eyes to become brittle, and
the likelihood of secondary exposure is high. As with other
intense light source exposure, you may experience a temporary
reduction in visual acuity, pain, discomfort, or visual impairment,
all of which increase the risk of secondary accidents.
▪......
... D isclaimer for risks involving patients with epilepsy.
▪......
... T he Medit i600 should not be used on patients that have been
diagnosed with epilepsy due to the risk of seizures and injury.
For the same reason, dental staff who have been diagnosed
with epilepsy should not operate the Medit i600.
5.7 Explosion Hazards
WARNING
▪......
... T he i600 system is not designed to be used near flammable
liquids or gases, or in environments with high oxygen
concentrations.
▪......
... T here is a risk of explosion if you use the i600 system near
flammable anesthetics.
5.8 Pacemaker and ICD Interference Risk
WARNING
▪......
... D o not use the i600 system on patients with pacemakers and ICD
devices.
▪......
... C heck each manufacturer's instructions for interference by
peripheral devices, such as computers used with the i600 system.
Intraoral Scanner i600
17