en
Issue
Saw blade not moving
Insufficient cutting per-
formance of the saw blade
The saw blade cannot be
coupled
Saw blade cannot be
uncoupled
10. Technical Service
Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
WARNING
►
Do not perform any servicing or maintenance
work under any circumstances while the product
is being used on a patient.
►
Do not modify the product.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
►
For
service
and
repairs,
B. Braun/Aesculapagency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
12
Cause
Loud running noise from motor Defective gearings
Saw blade teeth worn
Battery too weak
Battery worn/defective
Saw blade overheated
Coupling components or saw
blades deformed
-
Locking pin not fully releasing
the saw blade
please
contact
your
Detection
Blunt saw blade
Insufficient motor power/speed
Inadequate chip removal
Deformation
Incompatible saw blade
Button for saw blade release not pressed
completely
11. Accessories/Spare parts
Art. no.
GA675
GA676
GA678
GA679
GB243R
GB244R
national
GB600
GB495R
GB487R
GB488R
GB489R
TA014542
TA014543
Remedy
Have repaired by manufacturer
Replace saw blade
Charge battery
Replace battery
Move the saw back and forth during sawing and
ensure sufficient irrigation of the saw blade
Have repaired by manufacturer
Insert a compatible Aesculap saw blade, see
Accessories/Spare parts
Fully press the saw blade release button, see
Attaching and detaching the saw blade
Designation
Lid
NiMH battery long
Sterile funnel
Battery removal device
ECCOS sieve basket with fixation for Acculan 4
STERILIT Power Systems oil spray
ECCOS fixation for machine
ECCOS fixation for lid
ECCOS fixation for sterile funnel
ECCOS fixation for battery removal device
Instructions for use for oscillating saw GA331 (A4
for ring binder)
Instructions for use for oscillating saw GA331
(flyer)