Guidance And Manufacturer's Declaration - Electromagnetic Emissions - Covidien A-V Impulse 6060 Manual De Funcionamiento

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

The A-V Impulse 6000 Series Controller is intended for use in the electromagnetic environment specified below. The customer or the
user of the A-V Impulse 6000 Series Controller should assure that it is used in such an environment.
Immunity test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the A-V Impulse 6000
Series Controller is used exceeds the applicable RF compliance level above, the A-V Impulse 6000 Series Controller should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the A-V Impulse
6000 Series Controller.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distance between portable and mobile RF communications equipment and the A-V Impulse 6000 Series
Controller @ 3Vrms
The A-V Impulse 6000 Series Controller is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the A-V Impulse 6000 Series Controller can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the A-V Impulse 6000 Series Controller as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
EN-16

Guidance and manufacturer's declaration - electromagnetic emissions

IEC 60601
Compliance
Test level
3 Vrms
3 V/m
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
3.5
d =
v
1
0.12
0.37
1.17
3.69
11.67
Electromagnetic environment - guidance
Level
Portable and mobile RF communications equipment should be used no
closer to any part of the A-V Impulse 6000 Series Controller, including
cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = [1.17] √P
d = [1.17] √P 80 MHz to 800 MHz
d = [2.33] √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
each frequency range.
Interference may occur in the vicinity of equipment marked with the
following symbol:
80 MHz to 800 MHz
√P
d =
a
should be less than the compliance level in
b
3.5
√P
E
1
0.12
0.37
1.17
3.69
11.67
800 MHz to 2.5 GHz
7
√P
d =
E
1
0.23
0.74
2.33
7.38
23.33
A-V Impulse System
Tabla de contenido
loading