Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the user of
the VisionAire should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electromagnetic
environment – guid-
±6 kV contact
ance
±8 kV air
IEC 61000-4-2
±2 kV for power supply
lines
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for input/output
lines
±1 kV line(s) to line(s)
Surge
IEC 61000-4-5
±2 kV line(s) to earth
<5 % U
(>95 % dip in U
for 0,5 cycle
40 % U
(60 % dip in U
Voltage dips, short
interruptions and
for 5 cycles
voltage variations
on power supply
70 % U
input lines IEC
61000-4-11
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
magnetic field IEC
3 A / m
61000-4-8
NOTE U
is the A.C. mains voltage prior to application of the test level.
T
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
N/A
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5 % U
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
<5 % U
T
)
(>95 % dip in U
T
for 5 sec
3 A / m
Electromagnetic environment – guid-
ance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
T
)
T
T
Mains power quality should be that of a
)
typical commercial or hospital environ-
T
ment. If the user of the VisionAire re-
quires continued operation during power
mains interruptions, it is recommended
T
that the VisionAire is powered from an
)
uninterruptible power supply (UPS).
T
T
)
T
Power frequency magnetic fields should
be at levels characteristic of a typical lo-
cation in a typical commercial or hospital
environment.
PN MN238-C4 E | User Manual US - 21
VisionAire 5