Specifications - WIKA E-10 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

9. Specifications

9. Specifications
Specifications
EN
Measuring range
Overpressure limit
bar
psi
bar abs., psi abs.
For oxygen version
Vacuum resistance
Non-linearity per BFSL per
IEC 61298-2
Accuracy
Max. measuring deviation per
IEC 61298-2
Non-repeatability per IEC 61298-2
Temperature error at 0 ... 80 °C [32 ... 176 °F]
Zero point
Span
Long-term stability per DIN 16086
Reference conditions
Output signal
22
→ See product label
Measuring ranges ≤ 16 bar
Measuring ranges ≥ 25 bar
Measuring range 1,000 bar
Measuring ranges ≤ 300 psi
Measuring range ≥ 500 psi
Measuring ranges 600, 1,000,
1,500, 10,000 psi
Measuring range 15,000 psi
3.5 times
2 times
Yes
≤ 0.2 % of span
→ See "Max. measuring deviation per IEC 61298-2"
0.5 % of span
≤ 0.1 % of span
≤ 0.2 % of span/10 K
≤ 0.2 % of span/10 K
≤ 0.2 % of span/year
For use in hydrogen applications, observe the Technical information IN 00.40 at www.wika.com
regarding long-term stability.
Per IEC 61298-1
→ See product label
WIKA operating instructions pressure transmitter, models E-10 and E-11
3 times
2 times
1.7 times
3 times
2 times
1.7 times
1.43 times
Tabla de contenido
loading

Este manual también es adecuado para:

E-11

Tabla de contenido