5. Commissioning, operation
Sealing variants
Parallel threads
Seal the sealing face with flat gasket, lens-type
EN
sealing ring or WIKA profile sealing.
Tapered threads
Wrap threads with sealing material
(e.g. PTFE tape).
Mounting the instrument
The max. torque depends on the mounting point (e.g. material and shape). If you have any questions,
please contact our application consultant.
→ For contact details, see chapter 1 "General information" or the back page of the operating instructions.
1. Seal the sealing face (→ see "Sealing variants").
2
At the mounting point, screw the pressure transmitter in hand-tight.
.
3
Tighten the process connection hexagon with a torque spanner using the spanner flats.
.
Max. torque: 50 Nm [36.9 foot-pound]
Please observe that the threads are not skewed.
14
per EN 837
NPT, R and PT
WIKA operating instructions pressure transmitter, models E-10 and E-11
per ISO 1179-2 (formerly
DIN 3852-E)