2. Slide the insert into the shaft tube.
3. Screw the insert onto the joint sleeve (2) as far is it goes.
4. Press the rear release button (1) on the handle and slide the
shaft tube as far as it goes and latch into the handle.
6.4

Connecting the product

Note: On a VIO 300 D, the BiClamp upgrade must be installed.
These products possess an electronic identifier (i.e. the product
and its specifications are recognized by the unit after the con-
necting cable has been inserted, and several default settings are
executed automatically on the unit).
1. VIO 300 D: Connect the product to an MF-2 socket.
VIO 3: Connect the product to an MF-U socket.
2. Check whether the instrument has been detected: On the
VIO 300 D, "BiClamp" mode is displayed and on the VIO 3,
the "BiClamp" instrument is shown.
If a correctly connected instrument is not detected, replace
the instrument.
If the error persists, replace the electrosurgical unit.