Cleaning
1. Make sure the instrument jaws are open.
2. Immerse the product completely in a cleaning bath, ensuring
that the product does not come into contact with other com-
ponents in the bath.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Move the movable parts back and forth (at least 5 times).
5. Brush the inside of the shaft tube with a round brush (at
least 5 times).
6. Rinse the surfaces with sterile, deionized water (at least
1 minute).
7. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
8. Repeat the preceding cleaning steps if the product or used
rinse water still contains visible contamination.
Disinfection
1. Make sure the instrument jaws are open.
2. Immerse the product completely in a disinfectant bath, en-
suring that the product does not come into contact with oth-
er components in the bath.
3. Move the movable parts back and forth (at least 5 times).
4. Completely fill a disposable syringe with disinfecting solution
and rinse the lumina (at least 5 times).
5. Rinse the surfaces with sterile, deionized water (at least
1 minute).
6. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
7. Disconnect the irrigation tube from the shaft tube.
8. Dry the product, including the lumens, until there are no
more fluid residues visible (e.g. with filtered compressed air).