Medela Dominant Flex Instrucciones De Uso página 389

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 306
Electromagnetic Compatibility (EMC, IEC 60601-1-2, Table 6)
Recommended separation distance between portable and
mobile RF communications equipment and Dominant Flex
Dominant Flex is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled� The customer or the user of Dominant Flex can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and Dominant Flex as
recommended below, according to the maximum output power of the communications
equipment�
Rated maximum
output power
of transmitter
W
0�01
0�1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer�
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies�
NOTE 2 These guidelines may not apply in all situations� Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people�
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 0.35 √ P
0�04
0�11
0�35
1�11
3�5
80 MHz to 800 MHz
d = 0.35 √ P
0�04
0�11
0�35
1�11
3�5
Technical documentation
800 MHz to 2.5 GHz
d = 0.7 √ P
0�07
0�22
0�7
2�21
7