System Maintenance Work; Cleaning; Fuses; Transport - HEBU medical HBS x-touch 300 argon BiSeal Instrucciones Para El Uso

Aparato de cirugía electrodoméstica
Ocultar thumbs Ver también para HBS x-touch 300 argon BiSeal:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

HBS x-touch 300 argon BiSeal
10

System maintenance work

10.1 Cleaning

The electrosurgical unit can be wiped using a cloth dampened with standard disinfectant, and so kept
clean using the simplest possible method. Avoid allowing cleaning fluid and disinfectant to penetrate
into the unit.

10.2 Fuses

The fuse sockets are located on the back of the unit. The unit works with two fuses of 6.3 A each.
Exchanging the fuses:
Disconnect the unit from the mains
Release the fuse from the retainer using a flat blade screwdriver
Exchange the fuse
After an exchange of fuses, the unit must be checked by suitably qualified personnel.

10.3 Transport

Prevent mechanical damage and moisture.
If the unit has been stored for a long period in a cold environment, wait before removing the protective
packaging until the unit has reached room temperature.
During transportation, the standard safety measures additionally apply.
Only ever transport the unit in its original packaging.

10.4 Environmental directives

From the implementation date of the European directive 2002/96/EU as
national legislation, the following regulations apply:
Electrical and electronic units must not be disposed of with normal
household waste.
The consumer is required by law to return electrical and electronic devices
at the end of their service life to designated public collection points or to the
place of purchase.
Disposal details are set out by the applicable local legislation in the relevant
country. The symbol on the product, the instructions for use or the packaging
indicates the relevant applicable legislation. By sending end-of-life devices for
reuse, material recycling or other forms of utilization you are making a major
contribution to environmental protection.
Please also send the unit packaging for environmentally responsible recycling
after the end of the product's life.
®