NV: Not Valid.
If the measured value is exceeded, OV (overrange) is sent.
The associated values are sent on 1 line separated by a space (ASCII 32).
3.2.2.38
Setup.AutoInfo.Status
Setup.AutoInfo.P
Setup.AutoInfo.T.G
Setup.AutoInfo.T.R
Setup.AutoInfo.T.S
Setup.AutoInfo.T.B
Setup.AutoInfo.T.F
Setup.AutoInfo.T.E
Setup.AutoInfo.I
Setup.AutoInfo.O
Automatic message as soon as a change appears.
.Status:
Switches all set AutoInfos on/off.
.P:
PowerOn: Simulation PowerOn has been performed (3.2.2.39). No message
on power on.
.T.G:
Go: Automatic determination has been started.
.T.R:
Ready: Automatic determination has been ended.
.T.S:
Stop: An automatic determination has been stopped.
.T.B:
Begin: In the automatic determination, the sample heating begins.
.T.F:
Final: End of the sample heating in the automatic determination.
.T.E:
Error: Message together with error number.
.I:
Change in an input line of the "Remote" socket.
.O:
Change in an output line of the "Remote" socket.
Each message is sent with space (ASCII 32) and ! as introducers. The name of the device is
then sent (3.2.2.13). Special ASCII characters in the device name are ignored. If no device
name has been entered, only ! is sent.
This is followed by information regarding the node which initiated the message. In the case
of errors, the error number is sent.
Examples:
!Otto".T.G"
Device Otto has been started.
!".T.E;E26"
Device without a name has error E26.
3.2.2.39
Setup.PowerOn
Simulation of power on. The device has the same status as after power on: The boat is in
position 0 and the sample number on 0.
3.2.2.40
Setup.Initialise
Setup.Initialise.Select
Sets default values for the following areas:
Mode:
Branch Mode.
Config:
Branch Config.
All:
Values of the entire tree.
Setup:
Branch Setup.
Assembly:
Branch Assembly.
The action is initiated with &Setup.Initialise $G.
768 KF Oven
3.2 Remote control commands
Mode, Config, All, Setup, Assembly
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
$G
$G