Metrohm 768 KF Manual Del Usuario página 28

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

3.2.2.17
Info.Report
Output of the selected report.
3.2.2.18
Info.Report.Select
Selection of the report.
configuration: Configuration report. Not accessible when a determination is running.
parameters:
Parameter report. Only "live" parameters when a determination is running.
result:
Result report of the last completed determination.
The report output must be triggered with &Info.Report $G, see 3.2.2.17.
Reports which are sent automatically by the KF Oven start with a space (ASCII 32) and '.
This is followed by the individual identification for each report. Reports which are requested
via RS232 ($G) have the same individual identification of the report blocks, but are started
with ' only (no preceding space).
3.2.2.19
Info.Results.PurgeTime
Info.Results.CondTime
Info.Results.SampleHeatTime read only
Info.Results.LowTemp
Info.Results.HighTemp
Info.Results.GasFlow
Info.Results.LowFlow
Info.Results.HighFlow
Inquiry of the current results.
.PurgeTime: Purge time.
.CondTime:
Conditioning time.
.SampleHeatTime: Time during which the sample has been heated.
.LowTemp:
Lowest temperature during the sample heating.
.HighTemp:
Highest temperature during the sample heating.
.GasFlow:
Mean gas flow during the sample heating.
.LowFlow:
Lowest gas flow during the sample heating.
.HighFlow:
Highest gas flow during the sample heating.
3.2.2.20
Info.ActualInfo.Inputs.Status
Info.ActualInfo.Inputs.Change read only
Info.ActualInfo.Inputs.Clear
Info.ActualInfo.Outputs.Statusread only
Info.ActualInfo.Outputs.Changeread only
Info.ActualInfo.Outputs.Clear
Status sends the current status of the I/O lines. Change sends information regarding
whether a change in the status of a line has taken place since it was last cleared, Clear
clears the byte of Change.
For the output, the byte is converted to binary, e.g.
|0|0|0|0|1|0|1|0|
Line No.
|7|6|5|4|3|2|1|0|
1 means ON or change; 0 means OFF or no change.
The lines are assigned as follows (see also pages 53, 54):
Inputs:
0 Start (Pin 21)
1 Stop (Pin 9)
2 Terminate (Pin 22) 2 Stop (Pin 4)
3 Pin 10
4 Pin 23
5 Pin 11
6 Pin 24
7 Cond.ok (Pin 12)
768 KF Oven
Output: 2 1 + 2 3 = "10"
Outputs:
0 Ready (Pin 5)
1 Start (Pin 18)
3 HeatSmpl (Pin 17)
4 Terminate (Pin 3)
5 Error (Pin 16)
6 Pin 1
7 Pin 2
3.2 Remote control commands
configuration, parameters, result
$G
$G
read only
read only
read only
read only
read only
read only
read only
read only
$G
25
Tabla de contenido
loading

Tabla de contenido