3.1 General rules3.1 General rules
3.1.3 Status and error messages
Detailed status conditions
Status conditions of the global $G:
$G.Mode .Inac
.PurgeTime
.CondTime
.HeatSmpl
.Terminate
$G.Assembly .Prep.Wait
.Boat
Status conditions of the global $R:
$R.Mode .Ready
$R.Assembly.Ready
Status conditions of the global $S:
The instrument gives the status from which it has been stopped. The detailed status information is
therefore identical as for the global status $G.
Error messages
E36
E37
E38
E39
18
Waiting during start delay
Waiting during purge time
Waiting during cond.time
Heating the sample
Carrying out the terminating steps
Heating the oven to the set temperature
The sample boat has been manually moved
Inactive: Ready to start an automatic determination
An Assembly step has been carried out. Ready to carry out
another Assembly step (&Mode $G will trigger error message
E31).
RS receive errors:
Parity error
Exit: <QUIT> and set parity at both devices the same.
Stop Bit
Exit: <QUIT> and set stop bit at both devices the same.
Overrun error. At least 1 character could not be read.
Exit: <QUIT>.
The internal receive buffer is full (>82 characters).
Exit: <QUIT>.
768 KF Oven