3.2 Remote control commands3.2 Remote control commands
3.2.2.21
Info.ActualInfo.Meas.CyclNo
Info.ActualInfo.Meas.SampleTemp
Info.ActualInfo.Meas.OvenTempread only
Info.ActualInfo.Meas.GasFlowread only
Inquiry of the current values.
NV: Not Valid.
If the measured value is exceeded, OV (overrange) is sent.
A time frame can be generated from the cycle number and the cycle time (see 3.2.2.24).
The cycle number is always zeroed on start and on termination of an automatic
determination.
3.2.2.22
Info.ActualInfo.Status.BoatPosread only
Info.ActualInfo.Status.Valve read only
Info.ActualInfo.Status.Pump read only
Info.ActualInfo.Status.Heating read only
Status inquiry of the individual assemblies in the following format (examples):
.BoatPos
Position of the sample boat
.Valve
Position of the valve
.Pump
Status of the pump
.Heating
Heating level
3.2.2.23
Info.ActualInfo.Display.L1
Info.ActualInfo.Display.L2
1st and 2nd line of the display. The display can be written to.
The display is not operated by the KF oven if &Setup.Lock.Display is set to ON, see
3.2.2.35.
$Q sends the contents of the corresponding display line.
3.2.2.24
Info.Assembly.CycleTime
Inquiry of the cycle time in s.
3.2.2.25
Assembly.Prep
Switch heating to the sample temperature on/off.
3.2.2.26
Assembly.Heat
Assembly.Heat.Value
Control heating. The set heating power level is started with &Assembly.Heat $G.
0 means switch off heating.
3.2.2.27
Assembly.Valve
Assembly.Valve.Pos
Set valve. The selected position is set with &Assembly.Valve $G.
3.2.2.28
Assembly.Boat
Assembly.Boat.Rate
Assembly.Boat.Pos
Assembly.Boat.SetPos.InPos
Assembly.Boat.SetPos.OutPos0...130.0
The movements of the sample boat are initiated with &Assembly.Boat $G.
26
23 mm
"purge" or "transfer"
"ON" or "OFF"
20
up to 24 ASCII characters
up to 24 ASCII characters
read only
read only
read only
$G, $S
$G
0...50
$G
purge, transfer
$G, $S
0.1...10
0...130.0
0...130.0
768 KF Oven