768 KF Oven
RS send errors:
DSR=OFF. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
DCD=ON. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
CTS=OFF. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
The transmission has been interrupted with XOFF for at least 3
s.
Exit: Send XON or <QUIT>.
The RS parameters are no longer the same for both devices.
Reset.
The receive buffer contains an incomplete command (L F
missing). Sending from the instrument is therefore blocked.
Exit: Send L F or <QUIT>.
Device-specific errors:
Manual stop.
Exit: The error message disappears on the next start.
Check temperature sensor of sample temperature.
Exit: Rectify fault.
Sample temperature not OK.
Exit: The error message disappears when the fault is rectified
or &M $S.
Gas flow too low.
Exit: The error message disappears when the fault is rectified
or &M $S.
The attached titrator is not conditioned.
Exit: The error message disappears when the fault is rectified
or &M $S.
The oven temperature is higher than 360 °C.
Exit: The error message disappears when the fault is rectified.
The temperature sensor for the oven temperature is not OK.
Exit: Rectify fault.
The gas flow sensor is not OK (flow >500 mL/min).
Exit: Rectify fault. Continuation is possible only when the
minimum gas flow is set to 0 mL/min.