Metrohm 768 KF Manual Del Usuario página 22

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

E40
E41
E42
E43
E44
E45
E26
E135
E154
E163
E164
E165
E168
E169
768 KF Oven
RS send errors:
DSR=OFF. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
DCD=ON. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
CTS=OFF. No proper handshake for more than 1 s.
Exit: <QUIT>. Is the receiver switched on and ready to receive?
The transmission has been interrupted with XOFF for at least 3
s.
Exit: Send XON or <QUIT>.
The RS parameters are no longer the same for both devices.
Reset.
The receive buffer contains an incomplete command (L F
missing). Sending from the instrument is therefore blocked.
Exit: Send L F or <QUIT>.
Device-specific errors:
Manual stop.
Exit: The error message disappears on the next start.
Check temperature sensor of sample temperature.
Exit: Rectify fault.
Sample temperature not OK.
Exit: The error message disappears when the fault is rectified
or &M $S.
Gas flow too low.
Exit: The error message disappears when the fault is rectified
or &M $S.
The attached titrator is not conditioned.
Exit: The error message disappears when the fault is rectified
or &M $S.
The oven temperature is higher than 360 °C.
Exit: The error message disappears when the fault is rectified.
The temperature sensor for the oven temperature is not OK.
Exit: Rectify fault.
The gas flow sensor is not OK (flow >500 mL/min).
Exit: Rectify fault. Continuation is possible only when the
minimum gas flow is set to 0 mL/min.
3.1 General rules
19
Tabla de contenido
loading

Tabla de contenido