5.3
Replacement of Lamp
The instrument should be well cleaned before replacement of the lamp to prevent
dust from damaging the reflector. The replacement should be performed in a clean
environment.
Hot Surface
The lamp module becomes very warm due to the high operating
temperature. The lamp can also be damaged by fingerprints and
oily residues. Cotton gloves must be used when replacing the
lamp to avoid any damages or burn injuries.
Caution
Do not touch the lamp glass or reflector or let any rough surface
come into contact with the lamp glass. A microscopic scratch in
the glass might cause a lamp explosion later.
1.
Disconnect power supply to the instrument.
2.
Remove the two screws 1 fastening the protective guard from the lamp using a
3 mm allen key.
Fig. 5:5
Removing protective guard
1
Screw
5:6
Infratec™ Sofia
User Manual 6002 5350 / Rev. 6