5.2
Cleaning the Sample Handling Wheel (SHW)
If further cleaning is required, the four cells and SHW can be removed from the
instrument.
Caution
Do not move the SHW under any circumstances when the power
supply is connected to the instrument. This will damage the
motor connected to the SHW.
1.
Disconnect power supply to the instrument.
2.
Remove the drawer and remove/discard any grain kernels. If necessary, remove
any remaining threshing or dirt using the instrument cleaning brush.
Lift the instrument hood to expose the SHW. Unscrew the thumb screw 1 and
3.
remove the cover plate.
Fig. 5:2
Removing cover plate
1
Thumb screw
User Manual 6002 5350 / Rev. 6