Caution
Do not move the SHW under any circumstances when the power
supply is connected to the instrument. This will damage the
motor connected to the SHW.
Caution
Do not disconnect power to Infratec Sofia while the instrument
software is being updated.
Caution
Some force is required to remove the lamp module from the
instrument due to dust and water protection requirements. Take
care not to over-stretch the cable connecting the lamp to the
instrument.
Caution
The responsible body shall be made aware that, if the equipment
is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
Caution
If an ESD cap is provided, it should be mounted on any port not
being used.
Caution
Do not touch the lamp glass or reflector or let any rough surface
come into contact with the lamp glass. A microscopic scratch in
the glass might cause a lamp explosion later.
Caution
Infratec Sofia is to be connected to either 12 V DC (and NOT 24 V
DC) via Foss P/N 60024388, or to 100-240 V AC via a power adaptor
Foss P/N 60024086 (otherwise the CE-labelling of the instrument
can not be provided). Power adaptor may only be connected to
power outlet with protective earthing.
Hot Surface
The lamp module becomes very warm due to the high operating
temperature. The lamp can also be damaged by fingerprints and
oily residues. Cotton gloves must be used when replacing the
lamp to avoid any damages or burn injuries.
1:2
Infratec™ Sofia
User Manual 6002 5350 / Rev. 6