Performing a Control Test
You need the meter, a test strip, and control solution Level 1, Level 2, or both. The control level is
printed on the bottle label.
02/2012
1. Check the use by date on the
test strip container. Do not use
test strips past the use by
date.
2. Insert the test strip into the
meter in the direction of the
arrows. The meter turns on. A
test strip symbol and a
flashing blood drop symbol
appear on the display.
OR
3. Select the control solution to
test. You will enter the level
later in the test.
45