2. Start a tested program with the following properties:
– Sufficient cleaning effect.
– Thermal disinfection: 5 to 10 minutes at 90 to 95 °C,
A
≥ 3000.
0
– Final rinse with distilled or deionized water.
– Sufficient product drying (no visible fluid residues).
3. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
7.7

Check

1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
– Damage to the coating of the cryotip.
– Kinking and crimping of the application element.
If damaged, do not use this product!
7.8

Packaging

The product must be protected against damage during steriliza-
tion.
1. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/foil and/or in a ster-
ilization container.
28