Dentsply Sirona ProPex II Manual Del Usuario página 142

Ocultar thumbs Ver también para ProPex II:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 147
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
®
The ProPex
II is intended for use in the electromagnetic environment specified
below. The customer or the user of the ProPex
such an environment.
Immunity test
60601-1-2
Test level
Conducted
3 Vrms
disturbances
150 kHz to
induced by
80 MHz
RF fields
6 Vrms in
ISM bands
IEC
150 kHz to
61000-4-6
80 MHz
80% AM at
1 kHz
Radiated RF
10 V/m
IEC
80 MHz to
61000-4-3
2.7 GHz
46/48
IEC
Compliance
level
3 Vrms
150 kHz to
80 MHz
6 Vrms in
ISM bands
150 kHz to
80 MHz
80% AM at
1 kHz
10 V/m
®
II should ensure that it is used in
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the ProPex
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d = 1.17 √P
d = 1.17 √P
d = 2.3 √P
800 MHz to 2.7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
less than the compliance level in each
frequency range
Interference may occur in the vicinity of
equipment marked with the following
symbol:
B EN PIIL DFU WEB / Rev.07 / 02-2018 (Old ZF 190292.EN)
®
II, including
80 MHz to 800 MHz
a
, should be
b
.
Tabla de contenido
loading