Guidance and manufacturer's declaration - electromagnetic emissions
The Product is intended for use in the electromagnetic environment specified below.
The customer or the user of the Product should assure that it is used in such an
environment.
Emission test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic
Emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
36/48
Compliance
The Product uses RF energy only for its
internal function. Therefore, its RF emissions
Group 1
are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The Product is suitable for use in all
establishments, including domestic
Not applicable
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Not applicable
B EN PIIL DFU WEB / Rev.07 / 02-2018 (Old ZF 190292.EN)