3
M
AINTENANCE
3.1
Setting the LCD Contrast
The Ultima ID Pro™ features an adjustable LCD contrast for use in varying light conditions.
To adjust the contrast, press the 'SET' button which appears after the unit is powered on.
The screen will display several options as shown in (Figure 27). Pressing the 'SET' button
will display the options in (Figure 28). Press 'DOWN' or 'UP' to adjust the LCD contrast.
Press 'SAVE' to save your settings and return the (Figure 27).
TO FEED A
NEW PAPER ROLL
PRESS FEED
PRESS SET TO
ADJ LCD CONTRAST
BACK
3.2
Changing the White Plastic Sample Filter
When inspecting the sample filter, look completely around the entire outside diameter of the
white filter element located inside of the clear plastic housing. Look for red spots or the
beginnings of discoloration on the white outside diameter of the element. DO NOT look into
the round ends of the white element for red spots or discoloration. The round ends of the
filter may always appear red. If red spots or discolorations are discovered on the outside
diameter, the sample filter requires replacement to prevent the influx of particulate and oil
mists into the instrument.
To change the Sample Filter, first obtain a replacement filter, part number 6-02-6000-08-0.
Remove the existing filter from the retaining clip of the instrument by pulling straight up and
out. CAREFULLY remove the flexible, black rubber tubing connections from both ends of
the existing filter. DO NOT allow the tubes to slip back into the internal portion of the case.
Discard the existing filter in an environmentally friendly manner.
Install the tube ends onto the barbs of the replacement filter, taking note to align the flow
arrow of the filter with the flow arrow on the analyzers top panel. CAREFULLY slide the
tubing back into the internal portion of the instrument and seat the new filter into the retaining
clip. Inspect your hose assemblies for signs of oil entrapment. Replacement of the Sample
Filter usually requires replacement of the Brass Sample Hose Restrictor Assembly.
Manual Part Number:
5-06-7000-70-2
FEED
SET
Figure 27
& T
ROUBLESHOOTING
ADJUST CONTRAST
CURRENT VALUE
DEFAULT VALUE
DOWN
Manual File:
MN-A-0289 Rev. E
40
40
UP
SAVE
Figure 28
Page 3-23